Network Analyzer Mode (R&S ZVH-K42)
Performing Vector Measurements
Operating Manual 1309.6946.12 - 09
186
Calibrate the measurement for vector reflection measurements.
Vector measurements provide the full range of calibration methods for more accurate
results. Most full calibration methods require more than one calibration standard.
► Press the CAL key.
► Select the appropriate calibration (calibration on one or two ports).
► Perform calibration. For more information see "
".
► Reconnect the DUT.
The R&S ZVH shows the results of
the vector reflection measurement
on the filter.
You can change the measurement
configuration (e.g. the sweep time
or detector) or format (e.g. view
phase characteristics) without
affecting the accuracy of the
measurement. For more
information see "
" on page
Depending on the calibration method, you can also select another result display.