NRT
CONFig Menu
1080.9506.02
2.27
E-8
2.3.8
SPEC (Special Functions)
The
CONFig - SPEC
menu allows to set the reference value for relative measurements and the SWR
alarm mechanism.
CONFig - SPEC:
Selection
Description
Available
∆∆∆∆
REF
Reference value for measurement of relative
power variations in dB oder %
Always
SWR.ALARM
(De-)activating SWR monitoring
Sensor
IEC-Befehle:
see description of the menu items
∆
REF and SWR.ALARM
(Sections 2.3.8.1 and 2.3.8.2).
2.3.8.1
∆∆∆∆
REF (Reference Value for Relative Power Measurements)
The
∆
REF menu is used to define the reference value for indication of the result in the relative units %
or dB (see Section 2.3.9
)
.
CONFig - SPEC -
∆
REF:
The currently valid value is indicated upon calling up the menu. As an alternative to a user-defined ref-
erence value, the currently measured power can be used as a reference upon actuating the
RE F
key.
The reference value may be entered in W or dBm (use the cursor keys
and
for switchover at the
units position).
Unit:
W / dBm
Preset:
1 W
IEC/IEEE command:
[SENSe:]POWer:REFerence 0.0258 W
2.3.8.2
SWR.ALARM (SWR Monitoring)
The
CONFig - SPEC - SWR.ALARM
function is used to configure a mechanism for monitoring the SWR
of the load so that a warning can be output in the form of an acoustic or electrical signal. The menu item
SWR.ALARM described here is used to define the conditions under which an alarm is to be triggered.
The type of alarm has to be configured via the menu
UTIL-BEEPER
(acoustic, see section 2.5.7) or
UTIL-AUX/IO
(electrical, see section 2.5.9).
The alarm is triggered above a user-defined standing wave ratio
(SWR)
. To prevent the alarm mecha-
nism from responding at low uncritical powers, an additional threshold must be entered for the forward
power
(THRESHOLD),
below which no alarm will be given.
By entering a sufficiently high power threshold (1 MW or similar), the alarm mechanism can be totally
switched off. No acoustic alarm will be given for forward powers below this threshold, even if the maxi-
mum SWR is considerably exceeded.
After confirming the
SWR.ALARM
option in the
SPEC
menu, first the SWR-limit and then the power
threshold is requested. The entries must be confirmed with the SEL(LOCAL) key.
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...