:SYSTem:COMMunicate:SERial
This node provides the commands for checking the serial remote-control interface. The interface
is factory-set to 8 data bits, no parity and 1 stop bit. These settings cannot be changed. Regard-
ing the serial interface, the instrument is a DTE (data terminal equipment). The connection to the
controller must therefore be established via a modem bypass cable.
:SYSTem:COMMunicate[:RECeive]:BAUD 1200 | 2400 | 4800 | 9600 | 19200 | MINimum | MAXimum
| DEFault
The command defines the transmission rate for transmission and reception.
Example:
:SYST:COMM:SER:BAUD 1200
*RST value: none (last setting)
:SYSTem:COMMunicate:SERial[:RECeive]:PACE XON | NONE
This command switches the software handshake mode on (XON) or off (NONE).
Example:
:SYST:COMM:SER:PACE XON
*RST value: none (last setting)
:SYSTem:COMMunicate:SERial:CONTrol:RTS OFF | IBFULL | RFR
This command switches the hardware handshake mode off (OFF) or on (IBFull, RFR).
Example:
:SYST:COMM:SER:CONT:RTS OFF
*RST value: none (last setting)
:SYSTem:ERRor?
This query returns the next item from the error queue and removes that item from the queue.
Positive error numbers denote device-specific errors, negative error numbers denote the error
messages defined by SCPI (See Appendix B). If the error queue is empty, the NRT returns 0, ie
"No error". The operation is identical to that of STATus:QUEue[:NEXT]?.
Example:
:SYST:ERR?
Response: -221, “Settings conflict“
:SYSTem:PRESet
This command causes a reset of the instrument. It has the same effect as UTILity - SETUP -
RECALL 0 in the manual control mode or as the *RST command. This command triggers an
event and therefore has no *RST value.
Example:
:SYST:PRES
:SYSTem:VERSion?
This query returns the SCPI version number supported by the instrument This command is a
query and therefore has no *RST value.
Example:
:SYST:VERS?
Response: 1995.0
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...