SCPI-Specific Error Messages
NRT
1080.9506.02
B.2
E-3
Command Errors (contd.):
Error code
Error text in the case of queue poll
Error explanation
-109
Missing parameter
The command contains too few parameters.
Example: Command
:SENSe<n>:POWer:REFerence
requires input of a power.
-110
Command header error
The header of the command is faulty.
-111
Header separator error
The header contains an illegal separator.
Example: The header is nor followed by a "White space", "
ESE255
"
-112
Program mnemonic too long
The header contains more than 12 characters.
-113
Undefined header
The header is not defined for the instrument.
Example:
*XYZ
is undefined for every instrument.
-114
Header suffix out of range
The header contains an illegal numeric suffix.
Example: :
SENSe8
does not exist in the instrument.
-120
Numeric data error
The command contains a faulty numeric parameter.
-121
Invalid character in number
A number contains an invalid character.
Example: An "A" in a decimal number or a "9" in an octal number.
-123
Exponent too large
The absolute value of the exponent is larger than 32000.
-124
Too many digits
The number contains too many digits.
-128
Numeric data not allowed
The command contains a number which is not allowed at this position.
Example: Command
:SENSe<n>:FUNCtion:ON
requires specification of a text parameter.
-130
Suffix error
The command contains a faulty suffix.
-131
Invalid suffix
The suffix is invalid for this instrument.
Example: nHz is not defined.
-134
Suffix too long
The suffix contains more than 12 characters.
-138
Suffix not allowed
A suffix is not allowed for this command or at this position of the command.
Example: Command *RCL does not permit specifiing a suffix.
-140
Character data error
The command contains a faulty text parameter.
-141
Invalid character data
The text parameter either contains an invalid character or it is invalid for this command.
Example: Parameter input with write error
:SENSe<n>:BURSt:MODE USSER
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...