Status Reporting System
NRT
1080.9506.02
3.58
E-6
3.7.5
Reset Values of the Status Reporting Systems
Table 3-20 comprises the different commands and events causing the status reporting system to be
reset. None of the commands, except for *RST and SYSTem:PRESet influences the functional instru-
ment settings. In particular, DCL does not change the instrument settings.
Table 3-20
Resetting instrument functions
Event
Switching on
supply voltage
DCL,SDC
Power-On-Status-
Clear
(Device Clear,
Selected Device
Clear)
*RST or
SYSTem:PRESet
STATus:PRESet
*CLS
Effect
0
1
Clear STB,ESR
yes
yes
Clear SRE,ESE
yes
Clear PPE
yes
Clear EVENt parts of the
registers
yes
yes
Clear ENABle parts of all
OPERation-and QUES-
Tionable registers,
Fill ENABle parts of all
other registers with "1".
yes
yes
Fill PTRansition parts with
„1"
Clear NTRansition parts
yes
yes
Clear error queue
yes
yes
yes
Clear output buffer
yes
yes
yes
1)
1)
1)
Clear command process-
ing and input buffer
yes
yes
yes
1) Every command being the first in a command line, i.e. immediately following a <PROGRAM MESSAGE TERMINATOR>
clears the output buffer.