NRT
SENSe Subsystem
1080.9506.02
3.33
E-6
:SENSe<n>:FUNCtion:STATe? <Measurement Function>
This command returns the status of the measurement function:
Response 0: measurement function switched off,
Response 1: measurement function switched on.
All functions defined under :SENS<n>:FUNC:ON are available as measurement functions.
The command is a query and therefore has no *RST value.
Example:
:SENS2:FUNC:STAT?
"
POW:REV
"
Response: 1
:SENSe<n>:DATA? [<Measurement Function>]
This command provides access to the results of the current measurement(s). It is possible to re-
quest the result of a certain measurement function or to query the results of all measurement
functions switched on. The measurement functions are configured with the commands
:SENS<n>:FUNC....
The responses are output in the unit selected for the corresponding function. For some functions,
the unit can be selected. For more information, see Section 3.5.16, UNIT Subsystem.
By defining the measurement function, the result of a special measurement function is requested.
Without this specification the results of all active functions are queried. The responses are given
in the sequence of the measurement functions listed under [SENS<n>:]FUNC:ON.
The command is a query and therefore has no *RST value.
Example:
:SENS2:DATA?
"
POW:FORW:AVER
"
returns the measured result for the average forward power.
Response: 10
:SENSe<n>:INFormation?
This command returns the condensed data of the sensor in form of an ASCII string.
Example:
:SENS2:INF?
Response: sensor-dependent
:SENSe<n>:POWer
This node provides the following commands:
•
Setting of the integration time for NRT-Z sensors,
•
Entry of reference value for relative power indication (% or dB),
•
Entry of the threshold for the amplitude distribution function CCDF,
•
Setting of scale limits for bargraphs,
•
Configuration of rear-panel AUX TTL connector as output for monitoring power or matching
(but not both functions at a time).
Note:
If the AUX TLL connector is configured as an output, external triggering of the NRT
via this connector is not possible (see Section 3.5.15, TRIGger Subsystem).
:SENSe<n>:POWer:APERture 0.005s ... 0.111s | MINimum | MAXimum | DEFault
This command sets the integration time of the A/D-converters in sensors of the NRT-Z type (see
section 2.3.7).
Example:
:SENS1:POW:APER 0.1s
*RST value: DEF (sensor-dependent)
Note:
Specification of the unit (s) is optional.
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...