NRT
TEST Subsystem
1080.9506.02
3.43
E-6
:TEST:DIRect[?] "Sensor command"
This command is used for transferring direct setting commands to the active sensor. The setting
commands are contained in the sensor manual. With a ? appended, the command returns the re-
sponse of the sensor to the transferred command. This command has no *RST value.
Example:
:TEST:DIR
"
sensor command
"
:TEST:SENSor?
This command returns the status of the active sensor.
Example:
:TEST:SENS ?
The response is sensor-dependent and could read as follows:
"
NRT-Z44 V1.40
","
TEMPERATURE ERR
", "
CAL VALUES CHECKSUM ERR
"
The response always contains an identification string for the sensor with the type name and the
FW version number. In case of an error, error messages in plain text are appended. The content
of the response string is identical with the status query in manual control (see Section 2.5.11).