RS-232-C Interface
NRT
1080.9506.02
A.4
E-4
Annex A.2
RS-232-C Interface
The standard instrument is equipped with an RS-232-C interface. The 9-pin connector is located on the
rear panel. A controller can be connected via this interface for remote control.
Annex A.2.1
Interface characteristics
é
Serial data transmission in asynchronous mode,
é
Bidirectional data transmission via two separate lines,
é
Transmission rate selectable from 1200 to 9600 baud,
é
Logic 0 signal from +3 V to +15 V,
é
Logic 1 signal from -15 V to -3 V,
é
An external instrument (controller) can be connected,
é
Hardware handshake RTS/ CTS and software handshake XON / XOFF can be selected.
Fig. A-2 Pin assigment of the RS-232-C interface
Annex A.2.2
Signal lines
RxD (Receive Data),
Input; data line for transmitting from remote station to local terminal.
TxD (Transmit Data),
Output; data line for transmitting from local terminal to remote station.
DTR (Data Terminal Ready),
Output (log. ’0’ = active); with DTR, the instrument indicates that it is ready to receive data.
GND (Ground),
Interface ground, connected to instrument ground.
DSR (Data set ready),
Input (log. ’0’ = active); DSR indicates to the instrument that the remote station is ready to
receive data.
RTS (Request to send),
Output (log. ’0’ = active); with RTS, the instrument indicates that it is ready to receive data. The
RTS line controls whether the instrument is ready to receive data or not.
CTS (Clear to send),
Input (log. ’0’ = active); CTS tells the instrument that the remote station is ready to receive data.
RxD
DTR
1
6
5
9
TxD
DSR CTS
RTS
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...