1.4.5 Preselection of ON Period
To save further power, the NRT can be configured for automatic switch-off after a certain time in battery
mode. The NRT is automatically switched off if within a set period (5 min, 20 min or 2 h) neither a key is
pressed nor a remote-control command sent. The setting is made in the UTILity menu. In the following
example, automatic switch-off is activated after 20 min:
UTIL
Press the menu key.
Press the cursor key repeat-
edly until the automatic
switch-off menu is displayed.
Press the cursor key 2
×
to
shift the underscore to the
right. 20MIN is selected and
the underscore is blinking.
LOCAL
SEL
Press input key to confirm
the setting 20MIN. The NRT
returns to the measuring
mode.
Some seconds before switching off the NRT sends a repeated, rhythmical beep signal. If any key is
pressed during this time the switching off process is cancelled and the ON period is extended by the
selected value.
1.4.6 Battery Counter
The NRT contains two counters one of which registers the battery operating time since the last complete
charging of the battery. As a fully charged battery allows to operate the NRT for about 8 hours, this gives
an estimate of the remaining battery capacity. The battery time is displayed via the UTILity menu.
UTIL
Press menu key.
Press the cursor key repeat-
edly until the battery time is
displayed.
LOCAL
SEL
To return to the measuring
mode, press input key or any
key outside the menu field.
The battery counter is automatically reset after each complete charging of the battery (except for manual
stop of charging according to Section 1.4.3).
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...