NRT
SCPI-Specific Error Messages
1080.9506.02
B.7
E-3
Device Specific Error - sets bit 3 in the ESR register.
Error code
Error text in the case of queue poll
Error explanation
-300
Device-specific error
NRT-specific error not defined in more detail.
-310
System error
This error message suggests a device-dependent error. Please inform the R&S Service.
-311
Memory error
Error in the device’s memory.
-312
PUD memory lost
Loss of the protected user data stored using the *PUD command.
-313
Calibration memory lost
Loss of non-volatile calibration data used by the *CAL? command.
-314
Save/recall memory lost
Loss of the non-volatile data stored using *SAV? command.
-315
Configuration memory lost
Loss of the non-volatile configuration data stored by the instrument.
-330
Self-test failed
The selftest could not be executed.
-350
Queue overflow
This error code is entered into the queue instead of the actual error code if the queue is full. It indicates
that an error has occurred but not been accepted. The queue can accept 5 entries.
Query Error - Error in data request; sets bit 2 in the ESR register.
Error code
Error text in the case of queue poll
Error explanation
-400
Query error
General error occurring when data are requested by a query.
-410
Query INTERRUPTED
The query has been interrupted.
Example: After a query, the instrument receives new data before the response has been sent completely.
-420
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue operation.
-440
Query UNTERMINATED after indefinite response
A query is in the same command line after a query which requests an indefinite response.
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...