![Rohde & Schwarz NRT 1080.9506.02/.62 Operating Manual Download Page 149](http://html.mh-extra.com/html/rohde-and-schwarz/nrt-1080-9506-02-62/nrt-1080-9506-02-62_operating-manual_1477860149.webp)
NRT
Installation and Removal of Modules
1080.9506.02
4.11
E-4
Closing the in-
strument:
½
Proceed in reverse order to close the instrument. Press the two covers to-
gether when screwing on the rear-panel feet.
Attention:
Make sure that the braided cord is correctly fitted into the grooves of the case.
4.5.2 Option
NRT-B1
Installation:
½
Open the instrument (Section 4.5.1).
½
Cut out cover plate on rear panel using small side-cutting pliers (see Fig. 4-4).
NRT-B1
NRT-B3
Fig. 4-4
Removing the cover plates for options NRT-B1 and NRT-B3
½
Plug the option, with sub-D connector X1 pointing ahead, into the rear panel
from inside the instrument (see Fig. 4-5).
½
Fix the option with the two Phillips screws on the wiring side W1.
½
Tighten the two nuts of sub-D connector X1.
½
Connect cable W1 to the Mainboard (connector X1).
½
Close the instrument (Section 4.5.1).
Removal:
To remove the option, proceed in reverse order.
4.5.3 Option
NRT-B2
Installation:
½
Open the instrument (Section 4.5.1).
½
Remove tow round caps from the rear panel (see Fig. 4-5).
½
Insert panel jacks from outside into the rear panel.
½
Slide nut over the cable and tighten it.
½
Plug cable W8 from sensor 2 into connector X81 to X88 and cable W9 from
sensor 3 into connector X91 to X98 on the Mainboard. The correct position of
the connector is ensured by coding.
½
Set slide switch S1-2 to ON position (Mainboard).
½
Close the instrument (Section 4.5.1).
Removal:
To remove the option, proceed in reverse order.
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...