SCPI-Specific Error Messages
NRT
1080.9506.02
B.4
E-3
Execution Error - Error in executing the command; sets bit 4 in the ESR register
Error code
Error text in the case of queue poll
Error explanation
-200
Execution error
Error on execution of the command.
-201
Invalid while in local
The command is not executable while the device is in local due to a hard local control.
-202
Settings lost due to rtl
A setting associated with hard local control was lost when the device changed to LOCS from REMS or to
LWLS from RWLS.
-210
Trigger error
Error on triggering the device.
-211
Trigger ignored
The trigger (GET, *TRG or trigger signal) was ignored due to device timing considerations.
Example: The device was not ready to respond.
-212
Arm ignored
An arming signal was ignored by the instrument.
-213
Init ignored
Measurement initialization was ignored as another measurement was already in progress.
-214
Trigger deadlock
The trigger signal can not be used.
(The trigger source for the initiation of measurement is set to GET and the subsequent query is received.
The measurement cannot be started until a GET is received, but the GET would cause an interrupted-
error.)
-215
Arm deadlock
The arming signal can not be used.
-220
Parameter error
The command contains a faulty or invalid parameter.
-221
Settings conflict
There is a settings conflict between two parameters.
Example: FM1 and PM1 cannot be switched on at the same time.
-222
Data out of range
The parameter value is out of the range permitted by the instrument.
Example: Command *RCL only permits entries in the range of 0 to 4
-223
Too much data
The command is valid but contains too much data.
Example: The instrument does not have sufficient storage space.
-224
Illegal parameter value
The parameter value is invalid.
Example: An invalid text parameter is specified,
:INPut:PORT:POSition MIDDle
深圳德标仪器
135-1095-0799
Summary of Contents for NRT 1080.9506.02/.62
Page 2: ...深圳德标仪器 135 1095 0799 ...
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 6: ...深圳德标仪器 135 1095 0799 ...
Page 16: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 178: ...深圳德标仪器 135 1095 0799 ...
Page 190: ...深圳德标仪器 135 1095 0799 ...