7. SETUP SETTING MODE
IMR01N12-E7
47
Interval time (InT2)
This item is Interval time of Communication function (optional).
Data range:
0 to 250 ms
Factory set value:
10
Interval Time function:
The interval time for the HA430/930 should be set to provide a time for host computer
to finish sending all data including stop bit and switch the line to receive status for the
host.
If the Interval time between the two is too short, the HA430/930 may send data before the host
computer is ready to receive it. In this case, communication transmission cannot be conducted
correctly. For a successful communication sequence to occur, the HA430’s or HA930’s interval
time must match the specifications of the host computer.
Infrared communication address (Add3)
Infrared communication address is used to set the device address of the controller for Infrared communication
function.
Data range:
0 to 99
Factory set value:
0
Infrared communication speed (bPS3)
This item is communication speed of the Infrared communication function.
Data range:
9.6:
9600
bps
19.2: 19200 bps
Factory set value:
19.2
The factory set value of the Infrared communication speed of your PDA's
*
is 19200 bps.
Data bit, Stop bit and Parity bit on this controller under Infrared communication are fixed to the
following settings. The relevant bits on the PDA* are fixed to the same settings as on this controller
prior to factory set value. Therefore do not change them at the site.
Data bit: 7-bit
Parity bit: Even
Stop bit: 1-bit
* The PDA being used is necessary to be installed with the infrared communication software RKCIR.