No. Message Interpretation
Operator
action
920-922 Module
not
filled within
time limit in
wash with
cleaning
Module did not fill
properly with cleaning
solution.
- Check the solution
containers for adequate
volume. Replace as
required.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful Rinse
923 Inlet
not
refilled
properly
during bi-
directional
pump
calibration
The inlet module was
not filled properly
during bi-directional
pump calibration.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
924 Inlet's
liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 1
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
925 Inlet's
liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 2
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure
Removal
condition:Successful pump
calibration.
Continued on next page
11-61
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...