15. Radiometer settings
ABL800 FLEX Operator's Manual
15-18
Calibration verification
Regulations in some countries require verification of the reportable ranges selected
for the measured parameters, and a special feature described below is provided to
assist in the collection of this data.
Once the reportable ranges for the measured parameters are established, the limits
may be entered in the Reportable Ranges program described in
Analysis Setup
,
chapter 3
.
The procedure given below is only a suggestion to easily identify your specific
layout and measuring mode.
Step Action
1.
Make a new report layout called, e.g. "Cal. Verification", based on the
Radiometer default layout – see
Patient Report Setup
in
chapter 3
.
2.
Press
Patient ID layout
and deselect all items except for the Patient
ID and Sample type. Set default sample type to Cal.
Verification.
3.
Press
Edit Patient Results Layout
and deselect all parameters except
for the measured parameters. Deselect range checks for all parameters.
4.
In the
Analysis Setup - Syringe Mode Setup
dedicate one unoccupied
button to this specific mode (call it, e.g. Cal. Verification) using
Edit
name
.
Select desired parameters (press
Parameters
) and layout (press
Layout
) that you have made in step
1
above.
Continued on next page
Purpose
Preparatory
steps
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...