3. Installation and setup
ABL800 FLEX Operator's Manual
Quality control setup,
Continued
Rilibäk Ranges
(continued)
To activate or deactivate the Rilibäk Rules, do the following.
Step Action
Press the
On/Off
button to activate/deactiviate the assigned Rilibäk
rules.
1.
To add a new Rilibäk rule, do the following:
Step Action
1.
Press the
Add
button to display the screen below:
2.
Select the desired parameter from the parameter list shown in the right
side of the screen.
3.
Press the
Left
arrow button until the first "Lower Limit" box is
highlighted and enter the desired lower limit.
4.
Highlight the next box and select "<" or "<=".
5.
Highlight the first "Upper Limit" box and select "<" or "<=".
6.
Highlight the next "Upper Limit" box and enter the desired upper
limit.
Continued on next page
3-46