11. Troubleshooting
ABL800 FLEX Operator's Manual
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
869 Adaptive
sample
handling
applied
(decay to
pH/Oxi
without Bil)
870 Adaptive
sample
handling
applied
(decay to
pH/Oxi)
871 Adaptive
sample
handling
applied
(decay to
pH/BG/Oxi)
872 Adaptive
measuring
mode
applied
(pH/BG/Oxi
[2])
873 Adaptive
measuring
mode
applied
(pH/BG/
Met/Oxi)
874 Adaptive
measuring
mode
applied
(decay to all
micro
mode)
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
875 Sample
aged
The specified limit for
sample age has been
exceeded.
Draw and analyze new
sample.
Continued on next page
11-54
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...