ABL800 FLEX Operator's Manual
1. Introduction
Limitations of use and known interfering substances,
Continued
Substance Interference
Anticoagulants (sampling)
Anticoagulants that contain sodium salts will
give erroneously high
c
Na
+
results. Sodium
fluoride with or without EDTA and oxalate
(di Na) influence
c
Glu results. Sodium fluoride
gives erroneously high
c
Na
+
and low
c
Ca
2+
,
c
Glu and
c
Lac results. Tri sodium citrate
influences
c
Na
+
,
c
K
+
and
c
Glu results.
Thus Radiometer recommends the exclusive use
of heparin as anticoagulant. Solutions containing
organic preservatives may damage the ion-
selective membranes of the
c
K
+
and
c
Glu
electrodes when introduced into the analyzer.
Do not use EDTA, as it leads to erroneous pH,
p
CO
2
,
c
Na
+
,
c
K
+
and
c
Ca
2+
results. Use of
EDTA will also affect subsequent measurements
on the Ca electrode and it will reduce the
lifetime of this electrode.
Thiocyanic acid (degradation
product from treatment with
nitroprussid. Also produced in
thiosulphate treatment of cyanide
poisoning)
Erroneously high
c
Glu and
c
Lac measurements.
Glycolic acid (ethyleneglycol
degradation product)
Erroneously high
c
Lac measurements.
Insufficiently stabilized blood.
Caustic fluids (e.g. strong acids
or bases, detergents, etc.).
Fluids that precipitate.
Fluids that affect the sensor
enzymes.
Fluids that form complexes with
the analyzer solutions (calcium).
High viscosity fluids.
Hydrophobic fluids.
Reactive fluids.
Other fluids mode allows you to measure on
fluids other than heparinized human blood.
Be aware that some substances, such as listed in
the left column, measured in the Other fluids
mode may damage the instrument or the
electrodes. This can affect the subsequent
measurement on human blood or quality control
solutions.
CAUTION -
Known
interfering
substances
(continued)
Continued on next page
1-5
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...