11. Troubleshooting
ABL800 FLEX Operator's Manual
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
782 RADI-
ANCE
output
queue
cleared
The output queue was
cleared in RADIANCE
Connection Setup.
Informative message. No
action required.
783 Automatic
backup
started
Automatic backup
(selected in Disk
Functions Setup) has
started.
Informative message. No
action required.
785 Automatic
archiving
started
Automatic archiving
(selected in Disk
Functions Setup) has
started.
Informative message. No
action required.
786 Automatic
archiving
completed
Automatic archiving
(selected in Disk
Functions Setup)
completed successfully.
Informative message. No
action required.
787 Export
of
data logs
started
Export of data logs was
started by the user.
Informative message. No
action required.
788 Hypochlorite
detected -
remembrane
Glu, Lac, Cl
and both
Crea
electrodes
Measurement on a
substance resembling
hypochlorite has been
performed.
Any installed Glu, Lac,
Cl and Crea electrodes
have been damaged and
must be remembraned
before blood
measurements can be
performed.
As a precaution all
parameters are marked
with ‘Calibration
Expired’.
- Remembrane any
installed Glu, Lac, Cl and
Crea (A and B) electrodes.
- Perform a 2-point
calibration.
Removal condition:
Successful 2-point
calibration.
Continued on next page
11-48
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...