ABL800 FLEX Operator's Manual
4. Sample measurements
Measurements with FLEXQ,
Continued
An urgent
measurement
To bypass a queued measurement while the analyzer processes the samplers in the
sampler tray, do the following:
Step Action
1.
Check that
Urgent Manual Sample
is active on the
Ready
screen and
press it.
2.
If the urgent measurement violates parameter lifetime of one or more
queued samples, accept violation and book the inlet for manual
measurement by pressing
Yes
or abort booking of the inlet by pressing
No
.
3.
When allowed on the screen, open the inlet flap, insert the sampling
device in the inlet and press
Start
– see
Introducing a blood sample
in
this chapter.
Note that if the
Urgent Manual Sample
button is grayed out, the inlets are free and
the sample can be analyzed immediately.
Editing sample
registration
To change or add information to sample registration, do the following:
Step Action
Locate the required sample in the "FLEXQ Status" on the
Ready
screen and press the
button to proceed to the
Sample
Registration
screen.
1.
Change the patient data on the
Sample Registration
screen.
2.
Press
Accept
to confirm the patient data.
3.
4-11
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...