ABL800 FLEX Operator's Manual
9. Data management
Quality Control Log,
Continued
Filter/trend
function
(continued)
Step Action
5.
When all the criteria have been selected, press
Apply
to make the
search.
Press
Trend
to select the parameters.
6.
Select the desired parameter(s) by activating the check buttons (e.g.
pH,
p
CO
2
,
p
O
2
).
Press
View Trend.
7.
The trend of the selected parameter(s) is displayed.
To view the filter criteria, press the
View Filter
(the criteria cannot be
changed, only viewed).
To print the displayed trend, press
.
To return to the screen in step
5
, press
Back
.
Viewing quality
control statistics
Only those measurement values that fall within the statistical range are included in
the statistical data. Therefore parameters outside the measuring range or with a "
?"
are not included.
Parameter
•
The screen shows statistics for each control solution (identified by its slot,
type and lot number) for a single parameter.
Press
Next param.
. or
Prev. param.
. to view other parameters.
•
Press
WDC Report
to create a WDC Report – see
Creating a WDC Report
in
chapter 8
.
Continued on next page
9-13
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...