11. Troubleshooting
ABL800 FLEX Operator's Manual
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
589 Measured
QC value
higher than
statistical
range
The parameter value is
above the upper limit
of the user-defined
statistical range.
Measurement not
included into statistics.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
593 Insufficient
sample
Sample volume is too
small for the selected
measuring mode.
Affected parameters
will be marked with
"?".
- Make sure sample volume
matches the selected
measuring mode.
- Repeat the measurement,
ensuring sufficient sample
volume.
- Contact Radiometer
service representative.
597 Liquid
detected in
pH/BG
module
during Gas
Calibration
Unexpected liquid is
detected by the pH/BG
module during
calibration. Calibration
is aborted.
- Repeat calibration.
- Perform Fluid transport
troubleshooting procedure.
600
Demo
option will
expire soon
One or more currently
installed demo options
will expire and be
removed within the
next 7 days.
To have the option
permanently, contact a
Radiometer sales or service
representative.
Removal condition:
Options expired or
permanent options
installed.
604 Parameter
not installed
Parameter was not
installed or is
corrupted. Parameter
cannot be measured.
Contact Radiometer service
representative.
Continued on next page
11-28
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...