5. Quality control
ABL800 FLEX Operator's Manual
Quality control identification
Entering
information
with barcode
reader
To enter the information with the barcode reader, do the following:
Step Action
1.
Scan "Operator" from your ID-card.
If the name is included in the list of the analyzer's registered users, the
barcode will identify the person and fill in the name automatically.
Entering
information
manually
To enter information manually, do the following:
Step Action
1.
Highlight the item to be edited, using the arrow buttons.
Solution
and Lot
Are automatically detected and entered if only one quality
control system is used. Otherwise highlight the desired
solution from the screen and confirm with
Enter
.
If the same type of QUA/Au solution
is used, then only the solution type will be identified.
Choose the specific slot from the screen.
2.
For manual QC measurements only: Enter the temperature
(default: 25
°
C) on the keypad and confirm with
Enter
.
Mandatory entry is indicated by a
"
.
Continued on next page
5-8
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...