ABL800 FLEX Operator's Manual
2. What Is what
Screen elements,
Continued
Center section
The center section is the main information and interaction area of which there are
many different types.
The center section of the
Ready
screen is described part by part, starting from the
top (for the analyzers without FLEXQ module the center part contains the analyzer
type and – if entered in Miscellaneous Setup (see
chapter 3
) – analyzer message).
This part of the
Ready
screen contains the following elements:
•
Analyzer name
•
The messages to the operator regarding the FLEXQ activities
•
Urgent Manual Sample
request button. The button allows you to book an inlet
for an urgent measurement while the analyzer performs a task.
The button is grayed out when the analyzer is ready for any measurement.
This part of the
Ready
screen contains the following elements:
•
"Slot #": condition of each sampler tray slot (1, 2 and 3) is indicated with the
arrows that reflect the indicator diodes in the sampler tray:
Green
Place the sampler into the slot
or remove it from the slot.
Yellow
Do not remove the
sampler from the slot.
Blinking
green
Wait for the sampler to be
identified.
Blinking
yellow
Remove the sampler
from the slot.
•
"Sampler ID": appears when the sampler is placed in the sampler tray slot and is
registered. The name of this column can be changed in the Sample Logistics
Setup (see
Analysis Setup
in
chapter 3
of this manual).
•
"Time to Result": shows time left till a result will be ready; the clock indicates
that the sample age has exceeded settings selected in Sample Logistics setup.
•
"Status": gives information about a sampler in the slot:
Slot empty
A slot is ready to accept a sampler.
Ready for processing
Sampler has been registered in the slot.
Processing
Patient and sampler data are being processed.
Completed
Measurement is completed and the result can be
viewed.
Continued on next page
2-17
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...