6. Calibration
ABL800 FLEX Operator's Manual
General information
Purpose
The calibration process determines and checks the accuracy with which the
analyzer measures its parameters. The process is therefore important in ensuring
the reliability of results.
Calibrations are performed using solutions and gases of known concentration, for
each of the measured parameters.
Calibration
programs
The following calibrations performed on the analyzer are briefly described below.
(For detailed explanation please refer to the
Reference Manual, chapters 1,2,3.
)
Calibration Description
1 point
Measures each parameter on one solution and/or gas of known
composition, giving one value per parameter.
Relates the measured values to the theoretical values of a
solution and/or gas of the same composition.
Gives the Drift 1 and status values of the electrode.
2 point
Measures each parameter on two different solutions and/or
gases both of known composition, giving two values per
parameter.
Relates the measured values to the theoretical values of
solutions and/or gases of the same composition.
Gives the Drift 1, Drift 2 values and the status or zero point,
sensitivity of the electrodes (two sensitivity values are given
for the Crea B electrode and one for the Crea A electrode of
the ABL8x7 FLEX analyzers).
1 point pH/BG
(the USA only)
Measures
p
CO
2
and
p
O
2
on one gas mixture of known
composition and pH on a known solution, giving one value per
parameter
Relates the measured values to the theoretical values of a
mixture/solution of the same composition.
Gives the Drift 1 of the electrodes.
tHb
Calibrates the spectrophotometer. See
tHb calibration
in this
chapter.
Calibrations performed automatically at specific intervals selected in the
Calibration Schedule setup (see
Calibration Setup
in
chapter 3
) are called
scheduled
calibrations.
All calibration results are automatically recorded in the Calibration Log from
which they can be viewed and printed. Refer to
chapter 9: Data Management
for
details.
The calibration process and a detailed explanation of the terms are described in the
Reference Manual
.
Continued on next page
6-2
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...