5. Quality control
ABL800 FLEX Operator's Manual
Manual quality control measurement
Procedure
To make a manual measurement on a QUA or AutoCheck control
solution, do the following:
Step Action
1.
Open the syringe inlet flap.
2.
Place the adapter tip up into the syringe inlet.
3.
Press
Ampoule-QC
to select the measuring program.
4.
Press
Start
.
The sample is aspirated and the
Quality Control Identification
screen appears. Refer to the next section, for information on entering
data.
5.
When prompted by the analyzer, remove the adapter and close the
syringe inlet flap.
6.
Dispose of the used ampoules as infectious waste*.
*
Reference:
Clinical laboratory waste management. CLSI document GP5-A2.
High Crea
Check
S8377 Cleaning Met II Solution is used to check the analyzer performance at a
high creatinine level. The check up is made automatically during the system
alignment procedure.
To call an unscheduled High Crea Check, do one of the following:
•
Press
Menu > Start Programs > Run High Crea Check
•
On the
Analyzer Status
screen highlight the line for Slot 11 and press
Run
High Crea Check
to start a measurement.
5-6
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...