ABL800 FLEX Operator's Manual
15. Radiometer settings
15-7
Radiometer default settings,
Continued
Default settings for
Calibration Schedule
are as follows:
Activity Default
setting
2 Point Cal
8 hours
1 Point Cal
4 hours
1 Point pH/BG calibration
(for the USA only)
2 hours
tHb Cal
3 months
Cleaning 8
hours
Start time for 1- and 2-point calibration and
cleaning
00:00
Activity after measurement
None
Default settings for
Quality Control Statistics
are as follows:
Item Default
setting
Statistics factor
1.5
Remind to print statistics each month
No
Remind to export WDC data each month
No
Default settings for
Quality Control Mandatory Input
are as follows:
Item Default
Setting
Mandatory temperature
No
Default temperature
25
o
C
QC Schedule
settings are deleted.
AutoCheck Setup
: replacement warning before carousel empty: No.
Westgard Rules:
all rules are ‘Off’ and the use of Westgard Rules is disabled.
Rilibäk Ranges:
the use of Rilibäk rules is disabled.
Continued on next page
Calibration
Setup defaults
(continued)
Quality Control
Setup defaults
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...