ABL800 FLEX Operator's Manual
2. What Is what
Analyzer status,
Continued
Quality Control
status
The Quality Control status provides the following information:
•
status of the last measurement on each control solution type
•
messages referring to quality control measurements.
For the ABL8x7 FLEX analyzers, please note that Slot 11 will be automatically
installed when you scan the barcode for the Cleaning Met II Solution during
replacement.
Element Function
Solution
Lists the quality control slot and its solution type along
with a status symbol.
Cleaning Met II Solution (for ABL8x7 FLEX analyzers
only) is used for quality control of high creatinine levels.
OK
The last measurement was accepted.
?
One or more of the following occurred:
•
error in the last calibration.
•
analyzer error during last QC measurement.
•
a parameter measurement is outside the defined
ranges or a Westgard Rule has been violated.
The next measurement is overdue, and the
previous measurement, if any, was accepted.
?
The last quality control measurement had errors
present, and the next measurement is overdue.
Lot
The solution lot number for the slot.
Last Time
Lists the time the last measurement was performed.
Next Time
Lists the next scheduled time to perform the measurement
on the slot. See
Quality Control Schedule
,
chapter 3
.
Continued on next page
2-31
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...