ABL800 FLEX Operator's Manual
2. What Is what
Analyzer status,
Continued
Analyzer Status
elements
(continued)
Button
The colors indicate…
Quality Control
G
REEN
–
OK.
Status
Y
ELLOW
– error(s) in the last QC measurement and/or
QC schedule reminders.
RED –
error-prone parameters are repressed.
Reagents
G
REEN
–
no replacements due at the present time.
Status
Y
ELLOW
– a replacement is due, Calibration/Cleaning
solutions expire soon or Calibration/Cleaning
solutions have expired.
Electrode and
G
REEN
–
no replacements due at the present time.
others
Y
ELLOW
– a replacement is due.
System Messages
G
REEN
–
no (critical) messages.
Y
ELLOW
– non-critical
messages.
R
ED
–
critical messages. The analyzer cannot
calibrate or measure.
AutoCheck
G
REEN
–
no replacements due at the present time.
Status
Y
ELLOW
– a replacement is due.
Calibrations
status
The Calibrations status gives an overview of the status of the most recently
performed calibration of each type and relevant messages.
The following information is available on the screen:
Calibration Type
Lists each calibration type along with its status:
OK
Calibration was accepted.
?
Error(s) detected during calibration.
Pending or overdue calibration. The last
calibration was accepted.
Continued on next page
2-29
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...