ABL800 FLEX Operator's Manual
3. Installation and setup
Quality control setup
Programs
The quality control setup programs are described in this section.
Activate a button to enter the program.
QC solutions
This program allows you to assign or change a quality control solution to a specific
slot.
NOTICE:
Changing a control solution assigned to a slot will delete all current
quality control statistics obtained on that slot. If you want a copy of the statistics
for the last QC month, create a WDC Report – see
chapter 8
:
Disk Functions
.
NOTICE
: (for the ABL8x7 FLEX analyzers only) Slot 11 is assigned to the
Cleaning Met II Solution used for high Crea check.
Continued on next page
3-35
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...