ABL800 FLEX Operator's Manual
11. Troubleshooting
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
884 Sampler
tray
adjustment
needed
It was not possible to
position one or more
samplers directly below
the syringe inlet.
The affected sampler
tray positions in the
sampler tray must be
adjusted.
Contact your service
engineer to perform the
Sampler Tray Adjustment
procedure.
Removal condition: All
sampler tray positions are
adjusted.
885 Cyclic
QC
schedule
reset from
RADI-
ANCE
The cyclic QC schedule
has been reset and all
related reminders have
been removed as a
result of a RADIANCE
command.
No action required.
Informative message only.
887 Adaptive
measuring
mode
disabled
Due to insufficient or
inhomogeneous
sample, the analyzer
would normally
automatically have
chosen an adaptive
measuring mode but
this option has been
disabled.
Measurement is
aborted.
Repeat the measurement
with sufficient sample free
from clots and air bubbles.
888-889,
891
Pump
calibration
error
The pump calibration
failed for one or more
pumps. The previous
pump calibration is
used. A successful
pump calibration
should be attempted as
soon as possible.
- Repeat Pump calibration.
- Perform the Pump
Troubleshooting procedure.
Removal condition:
Successful pump
calibration.
Continued on next page
11-57
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...