5. Quality control
ABL800 FLEX Operator's Manual
Preparing a control solution
About control
solutions
Brief information is given below:
Item Description
Quantity
•
Each QC ampoule of QUA contains 2 mL
of QC solution.
•
Each QC ampoule of AutoCheck contains 0.7 mL of
QC solution.
Conditioning
before measure-
ment
Store the ampoules for five hours at 18-32
°
C. Control
solutions are sensitive to light, so always keep the ampoule
box closed.
Open ampoule
stability
In order to ensure the reliability of the measurement, each
QC ampoule must be used immediately after opening, for
one
measurement on
one
analyzer only.
Required
materials
The following materials are required to make a manual measurement on a control
solution:
Quality control system
Materials required
QUAL
QUAL
•
Four levels of control solutions
•
H700 Ampoule adapter
•
Ampoule opener
•
Rubber gloves
Aut
Aut
Aut
•
Four levels of control solutions
•
H700 Ampoule adapter
•
H705 Ampoule holder
•
Rubber gloves
Before measure-
ment on
QUALICHECK
+
control solution
Do the following:
Step Action
1.
Holding the ampoule between two fingers as shown,
shake it vigorously for at least 15 seconds.
2.
Tap the top of the ampoule until all of the solution collects at the
bottom.
Continued on next page
5-4
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...