3. Installation and setup
ABL800 FLEX Operator's Manual
3-88
Miscellaneous setup,
Continued
Analyzer
message
A message, sent from the RADIANCE system to the connected analyzer and
displayed on the
Ready
screen, can be changed or deleted in this program as
follows:
Step Action
1.
Press the keyboard icon to display the keyboard.
Type the message and confirm with
Enter
.
To delete the current message, press
Delete
on the keyboard, or delete
a message and type a new one if desired.
NOTICE:
The analyzer message can be up to 40 characters long.
Confirm the change with
Enter
on the keyboard to return to the
Miscellaneous Setup
screen.
2.
Screen saver
To set the time for the screen saver to appear, do the following:
Step Action
1.
Check that the screen saver check button is activated.
2.
In the "Screen saver" box select the desired time with the arrow
buttons.