ABL800 FLEX Operator's Manual
8. Disk functions
Saving setup
Purpose
You can copy your analyzer's current setup configuration onto a CD-RW,
removable drive, hard disk or network. It can be reloaded in the event the current
setup is lost or damaged or the same setup configuration should be loaded on other
analyzers without performing all the Setup programs.
Storing the
setup
To store the current analyzer setup, do the following:
Step Action
1.
Press
Edit Location
to select destination.
2.
Select required location by touching and highlighting it on the screen.
If a CD-RW is used, place a CD-RW in the CD-drive.
If a removable drive is used, connect it to the USB port.
Press
Expand/Collapse
to select the destination directory.
Press
Back
to return to the
Save Setup
screen.
3.
On the
Save Setup
screen press
Start
. When saving is complete, the
Ready
screen is displayed.
4.
Remove CD-RW (use the
Eject CD
button) or removable drive.
8-13
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...