ABL800 FLEX Operator's Manual
4. Sample measurements
Immediately before analysis
Purpose
It is necessary to mix a blood sample before introducing it into the system to
ensure its homogeneity. If a sample is transferred from a syringe to an analyzer
without being properly mixed, either the plasma phase or packed red blood cells
may be analyzed, rendering the oximetry results meaningless.
CAUTION - Risk of incorrect result
Always remove air bubbles from a sample before introducing it into the
analyzer. Air bubbles may result in erroneous
p
O
2
values.
CAUTION – Risk of erroneous results
Always meticulously follow the sampling procedures described in
chapter
12: Sampling
. Failure to follow these procedures may introduce clots or air
bubbles in the sample and yield erroneous results
.
Place a
safe
PICO sampler in the FLEXQ sampler tray. The sample will be
automatically mixed before the measurement.
FLEXQ module
Mixing a sample
For the analyzers without FLEXQ module, mix a blood sample in a sampling
device as follows:
Sampling device
Description
Mix a capillary sample with a magnet. Then move the
mixing wire to the end of the capillary opposite to that
from which the blood is to be aspirated.
Remove both capillary caps.
safe
PICO sampler or another syringe: Invert the syringe
repeatedly and roll it between the palms of your hands.
Invert the test tube repeatedly.
Then remove the cap.
CAUTION - Risk of incorrect result
Use capillary tubes with a volume relevant for the selected measuring
mode. Too small a volume will give the “Insufficient sample” error. Too
large a volume will render the internal analyzer corrections invalid and the
results inaccurate.
See
chapter 12
in this manual.
Guidelines for
handling
capillary blood
samples
4-9
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...