1026 Clot
(pH)
detection not
possible
The pH of the rinse
solution - stored
during the latest
calibration - was
invalidated when
either the Rinse
solution was
replaced, or when
maintenance was
performed on an
electrode.
A new pH value
needs to be
established by
performing a
calibration.
Perform a 1- or 2-point
calibration.
Removal condition(s):
Successful 1- or 2-point
calibration.
1027 Clot
suspected
beneath pH
electrode
The difference
between the pH of
the rinse solution,
measured during the
sample measurement,
and that stored after
the last calibration
exceeds the
maximum allowable
drift.
This indicates the
presence of a clot in
the measuring
chamber beneath the
pH electrode.
Remove the pH electrode and
check the measuring chamber for
clots: clean the measuring
chamber with a cotton stick
moistened with distilled water.
Make sure no cotton fibers are
left in the measuring chamber
Clean pH electrode
Remembrane Ref electrode
Replace pH electrode
Perform a 1- or 2-point calibration
Removal condition(s):
Acknowledge execution of the
"Clot removal procedure"
Successful 1- or 2-point
calibration
Step Action
1.
Press
Analyzer status > Electrodes and others.
2.
Press
Replace
.
Acknowledging
execution of a
"Clot-removal
procedure"
3.
Press
Clots removed
.
Data in this document will be added to the manual next time it is updated.
Radiometer Medical ApS
Åkandevej 21
2700 Brønshøj
Denmark
www.radiometer.com
Technical
documentation
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...