Communication Protocol
CubiScan 25 Command Set
CubiScan 25 Service Manual
6 8
Scale calibration codes are defined below.
Scrape Screen
This command causes the CubiScan 25 to save the current touchscreen image to the file
cs125isc.bmp.
Pos Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
6
Test Weight Value
Numeric
010.00-025.00
9
1
End of Text
Control
(ETX)
03h
10
1
Carriage Return
Control
(CR)
0Dh
11
1
Line Feed
Control
(LF)
0Ah
Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
1
Acknowledge
Alpha
(A)
41h
4
2
Identifier
Numeric
(00-99)
6
1
End of Text
Control
(ETX)
03h
7
1
Carriage Return
Control
(CR)
0Dh
8
1
Line Feed
Control
(LF)
0Ah
Negative Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
1
Neg. Acknowledge
Alpha
(N)
4Eh
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Pos Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scrape Command
Alpha
(a)
61h
3
1
End of Text
Control
(ETX)
03h
4
1
Carriage Return
Control
(CR)
0Dh
5
1
Line Feed
Control
(LF)
0Ah
Acknowledge Format