Communication Protocol
CubiScan 25 Command Set
CubiScan 25 Service Manual
6 7
Save Parameters
This command causes the CubiScan 25 to save all current parameters to the file param125.txt on
the SD card.
Scale Calibration
This function is required when the load cell is replaced, when a new controller is installed, or for
routine calibration. This command causes the CubiScan 25 to reply with a scale calibration code.
Each time this command is sent, the instrument prompts the operator to follow a defined
calibration pattern.
Scale calibration code descriptions:
•
SA01
Clear the CubiScan 25
•
SA02
Place test weight on the platform
•
SA03
Scale calibration complete
Pos Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
2
1
Save Command
Alpha
(s)
73h
3
1
End of Text
Control
(ETX)
03h
4
1
Carriage Return
Control
(CR)
0Dh
5
1
Line Feed
Control
(LF)
0Ah
Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Save Command
Alpha
(s)
73h
3
1
Acknowledge
Alpha
(A)
41h
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Negative Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Save Command
Alpha
(s)
73h
3
1
Neg. Acknowledge
Alpha
(N)
4Eh
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah