![Quantronix CUBISCAN 25 Service Manual Download Page 69](http://html.mh-extra.com/html/quantronix/cubiscan-25/cubiscan-25_service-manual_770138069.webp)
Communication Protocol
CubiScan 25 Command Set
CubiScan 25 Service Manual
6 0
Factor Toggle
This command is used to set the dimensional factor to either domestic or international.
Flash Firmware
This command causes the CubiScan 25 to flash file cs125.bin stored on the SD card.
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Pos Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
2
1
Fact. Toggle Command
Alpha
(F)
46h
3
1
Dom./Int'l
Alpha
(D/I)
44h or 49h
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Fact. Toggle Command
Alpha
(F)
46h
3
1
Acknowledge
Alpha
(A)
41h
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Negative Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Fact. Toggle Command
Alpha
(F)
46h
3
1
Neg. Acknowledge
Alpha
(N)
4Eh
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Pos Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
Pos Len
Description
Type
Range
ASCII