CubiScan 25 Service Manual
2 6
Chapter 6
Part Replacement
This section describes the procedures used to replace various parts of the CubiScan 25. The parts
are listed below.
•
•
“Replacing the Height Transmitter Board” on page 27
•
“Replacing the Height Receiver Board” on page 28
•
“Replacing the Width Transmitter Board” on page 29
•
“Replacing the Width Receiver Board” on page 31
•
“Replacing the Encoder” on page 32
•
“Replacing the Encoder Magnet” on page 34
•
“Replacing the Proximity Sensor” on page 35
•
“Replacing the Controller Box” on page 36
•
“Replacing the Motherboard” on page 38
•
“Replacing the Power Supply” on page 40
•
“Replacing the Scale Card” on page 42
•
“Replacing the Load Cell Summing Board” on page 43
•
“Replacing a Load Cell” on page 45
•
“Replacing/Adjusting Carriage Assemblies” on page 48
Replacing a Fuse
This section describes how to replace a fuse on the
CubiScan 25.
Items Needed
•
Replacement Fuse
Complete the following steps to replace a fuse.
1. Power the CubiScan 25 off.
2. Disconnect the power cord from the CubiScan 25.
3. The fuse drawer is located next to the power switch.
Open the fuse drawer by pushing the top and bottom
tabs simultaneously.
4. Remove the old fuse and replace it with the new fuse.
5. Shut the fuse drawer.