6.15 Alert Summary Table
The following table summarizes the alerts. See
Alarm and alert LED indicators
and
Alarm and alert audi-
ble indicators
for LED and audible indicators for each alert type. The screen below is an example of an alert
screen. The icons shown in the table below appear on the top of each alert screen.
Alert
Icon
Priority
Description
User Action
Data Activity: Do not
remove SD card.
Status
SD
card
read/write
underway.
No action needed.
Change Accepted
Status
Confirms acceptance
of prescription change
or device upgrade.
No action needed.
Oximetry: Good Connec-
tion
(icon only)
Status
Displays on the ther-
apy screen when the
blower is on and 3 sec-
onds of good connec-
tion is detected. Ap-
pears at the beginning
of therapy. This screen
will not display again if
the oximetry sensor is
removed and reapplied
unless therapy is
stopped and restarted.
No action needed.
Pair?: 123456
Yes/No
Status
Prompts to accept or
decline pairing to a
Bluetooth compatible
device. This device
can be identified by the
digits displayed.
Rotate control dial to accept
pairing (Yes), or decline
(No), then press control dial
to confirm selection.