Software (download and install from my.respironics.com):
• NI LabVIEW 2014 Support Package
• DreamStation Family Drivers, Version 1.0 or higher
• DreamStation Test Executive, Version 1.0 or higher (required if the DreamStation Field Service Applica-
tion is not already installed on your PC)
• DreamStation BiPAP autoSV, S/T and AVAPS Field Service Application (FSA), Version 1.0.1 or higher
9.2 Testing Prerequisites
Prior to testing any DreamStation device, the following must be ensured:
1. The device error log must be cleared.
2. The device real time clock must be accurate.
3. The DreamStation base unit and Humidifier must both be at room temperature. Refer to Section 9.3
for testing environmental specifications.
Refer to
Chapter 6
on using the Service Center Tools software to clear the device error log and to reset the
real time clock.
9.3 Testing Environment Specifications
The DreamStation devices must be tested within the following temperature specifications:
41 º F to 95 º F (+5 º C to 35º C)
9.4 Software Download and Installation
1. Download the
NI LabVIEW 2014 Support Package
from my.respironics.com, and install the software
by accepting all license agreements and default installation locations.
2. Download the
DreamStation Family Drivers
from my.respironics.com, and unzip the files to your
PC.
a. For Windows 7, run the installer and follow all prompts to install the drivers.
b. For Windows 10, refer to Section
9.4.1 Windows 10 DreamStation Family Driver Installa-
tion
.
3. Download the
DreamStation Test Executive
(if the DreamStation FSA is not already installed on
your PC), and install the Test Executive by accepting all license agreements and default installation
locations.
4. Download the
DreamStation BiPAP autoSV, S/T and AVAPS Field Service Application
(FSA)
from my.respironics.com, and install the software by accepting all license agreements and default in-
stallation locations.