P
AGE
66
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Code
Error Name
Error Level
Failed
Component
Actions
E-71
ERR_PSENS_STATUS_B
ITS_ERROR
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-72
ERR_PSENS_UNABLE_T
O_OBTAIN_BUS
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-73
ERR_SENSOR_PRESS_
OFFSET_STOP
STOP
PCA
1. Replace PCA and test.
E-74
ERR_PSENS_NO_CALLB
ACK
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and
test.
E-82
ERR_FLOW_SENSOR_O
FFSET
CONTINUE
PCA
1. Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
E-83
ERR_FSENS_UNABLE_T
O_OBTAIN_BUS
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-84
ERR_FLOW_SENSOR_S
TOP
STOP
PCA
1. Replace PCA and test.
E-85
ERR_FLOW_SENSOR_O
CCLUDED
CONTINUE
PCA
1. Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.