Removing/installing the Bottom/Flip Lid Assembly:
1. Remove the Tank Assembly, Flip Lid Seal, Dry Box Seal, and the Back Panel Assembly (refer to pre-
vious corresponding sections).
2. Install Back Panel Assembly, Flip Lid Seal, Dry Box Seal and Tank Assembly onto the new Bot-
tom/Flip Lid Assembly as instructed in the previous sections.
3. Place a new warning label and Serial/Model number label on the bottom of the Humidifier (refer to
Section 7.3
for creating the Serial/Model number label.
4. Route the heated tube wire harness through the Flip Lid opening as shown below.
5. Install ISO Port Cover over heated tube wire harness.