Device
Settings
:
Choosing this screen will take you to a sub-menu where you can adjust the way the device displays infor-
mation. These settings are described here.
Note:
Not all settings shown here will display on the device. The display will vary based on therapy device
model and device settings.
Icon
Text
Description
AHI
Show
AHI/Fit/PB
You can select whether or not the Apnea/Hypopnea index,
Mask Fit averages, and Periodic Breathing averages are
displayed on the Patient Info screens.
cmH
2
O
or
hPa
cmH
2
O
or
hPa
You can select the units of pressure that are displayed on the
screen.
Language
This feature allows you to choose which language to display
on the interface.
Clear Default
Reminders
This setting turns off the default patient reminders that are
enabled in the therapy device from the factory.
Note: This does not turn off additional reminders that you may
have activated in Encore. Encore messages must be cleared
or modified in Encore.
Reset Data
Use the Reset Data function to clear patient data from the ther-
apy device, as well as an SD card and modem (if installed).
After you press the control dial to execute Reset Data, the de-
vice will display a message asking you to confirm the reset.
Press the control dial again to reset data in the device.
Note: Reset Data resets Blower Hours that are visible to the
patient, but it does not reset Machine Hours in the Provider
Menu.
Reset Blower
Hours
Select Yes if you want to reset the blower hours (e.g., to track
device usage between patients).
Reset Therapy
Hours
Select Yes if you want to reset the therapy hours back to the
default of 0 hours.