Code
Error Name
Error Level
Failed
Component
Actions
E-191
ERR_ALS_BIST
CONTINUE
PCA
1. Clear error log and test.
2. If the retest fails or the
error reoccurs, replace the
PCA and retest.
E-192
ERR_USB_DEVICE_LOS
T
REBOOT
Accessory
Module
1. Remove and reinsert the
accessory module.
2. If error continues,
replace accessory module
if pulse oximetry is
required.
E-193
ERR_USB_HOST_NO_D
EVICE
CONTINUE
Accessory
Module
1. Remove and reinsert the
accessory module.
2. If error continues,
replace accessory module
if pulse oximetry is
required.
E-200
ERR_RTOS_ABORT_ER
ROR
STOP
PCA
1. Replace PCA and test.
E-201
ERR_RTOS_TIMEOUT
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and
test.
E-202
ERR_RTOS_ONE_SHOT
_TASK_NOT_STOPPED
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and
test.
E-210
ERR_PERF_CHECK_MS
G_TO
STOP
PCA
1. Replace PCA and test.
E-211
ERR_PERF_CHECK_PO
ST_FAIL
STOP
PCA