P
AGE
124
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Replacing the Back Panel Assembly
Kit
Included in Kit
Tools Required
Back Panel Assembly
•
Back Panel Assembly
•
# 4 x 1/2” screws (qty 5)
•
Torx Screwdriver (T10 is
recommended)
To remove the Back Panel Assembly:
1. Remove the Water Chamber Assembly (refer to previous corresponding section).
2. Pull the Back Panel away from the Humidifier while lifting up on the top to release the Back Panel tab
from the latch area.
To install the Back Panel Assembly:
1. Align the Back Panel onto the Bottom Assembly.
2. Secure the Back Panel with the four # 4 x 1/2” screws (torque to 6 IN-LB).
3. Verify the Back Panel is fully seated/secure.
Torque screws to 6 IN-LB
when installing Back Panel