Alert
Icon
Priority
Description
User Action
Pair?
Yes/No
Status
Prompts to accept or
decline pairing to a
Bluetooth compatible
device.
Rotate the control dial to
accept pairing (Yes), or
decline (No), then press the
control dial to confirm
selection. Selecting Yes
opens the Bluetooth
passkey screen.
Flow Sensor Error:
Limited Therapy
Notification
Flow
sensor
malfunction.
Press the control dial to
acknowledge and clear the
message. Disconnect the
device from power.
Immediately remove patient
from ventilator and connect
them to alternate source of
ventilation.
SD Card
Removed.
Notification
or Alert 2
Indicates SD card
has been removed
from therapy device
and not reinserted
before the start of the
current
therapy
session.
Reinsert SD card, or click to
clear alert.
Oximetry: Good
Study (icon only)
Notification
After exiting therapy,
the therapy device will
display the Good Oxi-
metry pop-up screen if
the oximetry sensor
maintained a good con-
nection for at least 4
consecutive hours
within a 24 hour win-
dow with breaks in ther-
apy no greater than 1
hour each.
Press Control Dial to
acknowledge and clear the
message.
SD Card Error:
Remove and Reinsert
Notification
SD card error
detected.
Remove SD card and reinsert.
If alert continues to occur,
replace with another card.
SD Card Full.
Notification
SD card is full.
Remove SD card and
replace with a new card.