P
AGE
74
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
6.6 Failure Mode Troubleshooting
The following table can be used as a guide to aid in troubleshooting the device based on the potential prob-
lem. All issues may not be presented here.
Problem/Failure Mode
Possible Cause
Steps to Take
Power Supply Related Error
•
Incorrect Power Supply
•
Power Cord not pushed
in the whole way
•
If returned, check to see if the
power supply is correct.
•
Make sure power cord is
pushed in the whole way.
No Power
•
All connections not
made.
•
Power Supply not
functioning
•
PCA Failure
•
Damaged PCA
components
•
Intermittent Connection
•
DC Power Cable not
pressed down/installed
the whole way
•
Check all connections.
•
Check Power Supply.
•
Check for damaged
components on PCA.
•
Check DC Power Cable
installation.
•
Replace any failed
components.
Airflow Does Not Turn On
•
Blower connections not
made.
•
Faulty Blower
•
Faulty PCA
•
Verify all connections are
made.
•
Replace Blower or PCA
(whichever is faulty).