P
AGE
67
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Code
Error Name
Error Level
Failed
Component
Actions
E-87
ERR_FLOW_SENSOR_B
US
CONTINUE
PCA
1. Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
E-93
ERR_RTC_VALUE
CONTINUE
N/A
1. Clear error log, set the
RTC and test.
2. If the test fails or the
error reoccurs, replace the
PCA and retest.
E-94
ERR_RTC_STOPPED
CONTINUE
PCA
1. Replace PCA and test.
E-100
ERR_HUMID_NO_HEAT
CONTINUE
Heater Plate
1. Clear error log and test.
2. Test humidifier
separately.
E-101
ERR_HUMID_MAX_TEM
P
CONTINUE
Heater Plate
1. Clear error log and test.
2. Test humidifier
separately.
E-105
ERR_HUMID_AMBIENT_
COMM
CONTINUE
PCA
1. Clear error log and test.
2. If the test fails or the
error reoccurs, replace the
PCA and retest.
E-106
ERR_HEATED_TUBE_M
AX_TEMP
CONTINUE
Heated Tube
1. Clear error log and test.
2. Test humidifier
separately.
E-107
ERR_HEATEDTUBE_DIS
CONNECT
CONTINUE
N/A
1. Clear error log and test.
E-108
ERR_HUMIDIFIER_DISC
ONNECT
CONTINUE
N/A
1. Clear error log and test.
E-130
ERR_TASK_WDOG_TIM
EOUT
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.